02-662 NTron Oxygen sensor, SenzTx-101 (0-25%) O2

03-908 Inmatec oxygen sensor alternatively, SenzTx-101 (0-25%) O2


The zirconia sensor offers fast response time and long service life with virtually no drift, whilst the electrochemical sensor allows measurement in background gases containing hydrocarbons.

Technical Data

Measurement technology Zirconia (ZR)
Measurement range 0 to 25%
Output resolution ( for %) 0.01%
Output resolution (for ppm) 1ppm
Accuracy +/-2% of reading (or 2ppm O2) @ calibrated temperature and pressure
Response time (T90) <10 seconds @ 25˚C (within selected range)
LDL (Sensitivity) 0.01% (when measuring %) / 1ppm (when measuring ppm)
Temperature range -20˚C to +50˚C
Pressure range 900 to 1100 mBarabs
Linearity +/- 2% of reading
Life expectation 3-5 years
Humidity  0-95% RH non-condensing
Shelf life No shelf life

Proven sensor technology 
With a choice of either zirconia or electrochemical sensor technology the SenzTx offers reliability, accuracy and flexibility. Both technologies have a broad measurement capability allowing the user to measure from selected ranges from 1ppm to 100% oxygen. 
Zirconia sensor 
The Ntron zirconia oxygen sensor is a nondepleting zirconia solid electrolyte sensor. A small capillary on the sensor controls the diffusion of oxygen into the sensor. When heated to over 400°C oxygen is electronically reduced causing current flow through the zirconia electrolyte. Zirconiumoxide allows the movement of oxygen ions through the substrate from a high to a low concentration. The measurement of oxygen is determined by the current flowing through the electrodes. The zirconia sensor has an unlimited shelf life without the loss of calibration and has an expected life of three to five years. The zirconia sensor is not position sensitive and has low cross sensitivity to other gases and does not dry out.

Monitoring oxygen levels via oxygen sensors

The Senz-TX oxygen sensors are designed to monitor oxygen levels in various applications, including compressed air systems. By accurately measuring oxygen levels, the Senz-TX sensor can help ensure that the compressed air quality meets the desired standards.
When using Senz-TX sensors for compressed air quality testing, it is important to consider the following factors:
  1. Accuracy: The Senz-TX sensor should be calibrated regularly to ensure accurate readings of oxygen levels in the compressed air. This will help in detecting any deviations from the desired oxygen concentration and maintaining the quality of the compressed air. 
  2. Monitoring: The Senz-TX sensor can be used to continuously monitor oxygen levels in the compressed air system. By tracking changes in oxygen concentration over time, operators can identify any issues that may affect the quality of the compressed air. 
  3. Maintenance: Regular maintenance of the Senz-TX sensor is essential to ensure its proper functioning. This includes cleaning the sensor, replacing any worn-out parts, and calibrating it as needed. 
  4. Integration: The Senz-TX sensor can be integrated into a larger monitoring system for comprehensive monitoring of compressed air quality. By connecting the sensor to a control system, operators can receive real-time data on oxygen levels and take immediate action if any deviations are detected.
Overall, the Senz-TX sensors can be effective tools for monitoring and maintaining compressed air quality. By following best practices for sensor usage and maintenance, operators can ensure that their compressed air systems meet the required quality standards.